簡介
C101
Series of small dimensions calibrators of AC and DC voltage and current up to 1100V and 100A, with high load possibility.
It allows checking of economical multimeters and clamps meter
The C101F Calibrator enables generating of:
Frequency of alternating signals may be chosen from values 50Hz (synchronized with power supply frequency), 60Hz, 400Hz or can be programmed in range 45.00...2000Hz. Output signal can be set by means of external keyboard or by standard interface RS232C from PC computer.
The C101F Calibrator is characterised by high accuracy, high output power, small dimensions and light weight.
Technical parameters of C101F Calibrator |
|||||||||||
Parameter |
Range |
||||||||||
100mV |
1V |
10V |
100V |
1000V |
1mA |
10mA |
100mA |
1A |
10A |
||
Resolution |
100nV |
1mV |
10mV |
100mV |
1mV |
1nA |
10nA |
100nA |
1mA |
10mA |
|
Settings range |
1... |
0.01... |
0.1... |
1... |
10... |
0.01... |
0.1... |
1... |
0.01... |
0.1... |
|
Uncertainty *) DC [% of set value+ |
0.03 |
0.02 |
0.02 |
0.03 |
|||||||
% setting range] |
AC |
0.05 |
0.05 |
0.05 |
|||||||
Maximum load |
20mA |
1100mA |
200mA |
30mA |
11V |
2V@20A |
|||||
Frequency |
DC; 50Hz synchronized with power supply; 60±0.01Hz and 400±0.1Hz 45.00...99.99±0.01Hz, 100.0...999.9±0,1Hz, 1000...2000±1Hz |
||||||||||
Temperature coefficient |
±0.1 of uncertainty value / 1°C |
||||||||||
Weight and dimensions |
16kg and 478 (width)x194 (height)x342 (depth) mm |
||||||||||
Power supply |
230V ±10% / 45…65Hz / 200VA |
||||||||||
DC – direct voltage and current, AC – alternating voltage and current |
The C101F Calibrator enables generating of:
Frequency of alternating signals may be chosen from values 50Hz (synchronized with power supply frequency), 60Hz, 400Hz or can be programmed in range 45.00...2000Hz. Output signal can be set by means of external keyboard or by standard interface RS232C from PC computer.
The C101F Calibrator is characterised by high accuracy, high output power, small dimensions and light weight.
Technical parameters of C101F Calibrator |
|||||||||||
Parameter |
Range |
||||||||||
100mV |
1V |
10V |
100V |
1000V |
1mA |
10mA |
100mA |
1A |
10A |
||
Resolution |
100nV |
1mV |
10mV |
100mV |
1mV |
1nA |
10nA |
100nA |
1mA |
10mA |
|
Settings range |
1... |
0.01... |
0.1... |
1... |
10... |
0.01... |
0.1... |
1... |
0.01... |
0.1... |
|
Uncertainty *) DC [% of set value+ |
0.03 |
0.02 |
0.02 |
0.03 |
|||||||
% setting range] |
AC |
0.05 |
0.05 |
0.05 |
|||||||
Maximum load |
20mA |
1100mA |
200mA |
30mA |
11V |
2V@20A |
|||||
Frequency |
DC; 50Hz synchronized with power supply; 60±0.01Hz and 400±0.1Hz 45.00...99.99±0.01Hz, 100.0...999.9±0,1Hz, 1000...2000±1Hz |
||||||||||
Temperature coefficient |
±0.1 of uncertainty value / 1°C |
||||||||||
Weight and dimensions |
16kg and 478 (width)x194 (height)x342 (depth) mm |
||||||||||
Power supply |
230V ±10% / 45…65Hz / 200VA |
||||||||||
DC – direct voltage and current, AC – alternating voltage and current |