簡介
FLUID 100
Operating range:-18 °C / +250 °C, (-0,4 °F / 482 °F)
Applications:
Check and calibration of temperature sensor in the laboratory as well as in the field in compliance of ISO 9000 regulations
Checking glass thermometers
Checking thermostats
Technical specification | FLUID 100 | FLUID H100 | FLUID 200 | FLUID H200 |
Operating range | -18 °C ÷ + 125 °C (*) (-0,4 °F ÷ 257 °F) | -18 °C ÷ + 140 °C (*) (-0,4 °F ÷ 284 °F) | Room temp. ÷ +200 °C Room temp. ÷ +392 °F | Room temp. ÷ +250 °C Room temp. ÷ +482 °F |
Average heating time (stabilization included) | 12 °C/min (53 °F/min) |
12 °C/min (53 °F/min) |
10 °C/min (50 °F/min) |
10 °C/min (50 °F/min) |
Average cooling time (stabilization included) | 5 °C/min (41 °F/min) |
5 °C/min (41 °F/min) |
4 °C/min (39 °F/min) |
4 °C/min (39 °F/min) |
Axial uniformity | ±0,05°C / ±0.09°F at 0°C / 32°F (for 60 mm from the bottom) |
±0,05°C / ±0.09°F at 0°C / 32°F (for 60 mm from the bottom) |
±0,05°C / ±0.09°F at 0°C / 32°F (for 60 mm from the bottom) |
±0,05°C / ±0.09°F at 0°C / 32°F (for 60 mm from the bottom) |
Radial uniformity (at 40 mm) | ±0,02 °C (at –5 °C) ±0.03 °F (at 23 °F) |
±0,02 °C (at –5 °C) ±0.03 °F (at 23 °F) |
±0,05 °C (at 50 mm) ±0.09 °F (at 50 mm) |
±0,05 °C (at 50 mm) ±0.09 °F (at 50 mm) |
Hole depth | 170 mm (6.6 in.) |
170 mm (6.6 in.) |
170 mm (6.6 in.) |
170 mm (6.6 in.) |
Hole diameter | 60mm (2.3 in.) | 60mm (2.3 in.) | 60mm (2.3 in.) | 60mm (2.3 in.) |
Display accuracy | ±0,15 °C (±0.27 °F) |
±0,15 °C (±0.27 °F) |
±0,15 °C (±0.27 °F) |
±0,2°C (±0.3 °F) |
Stability | ±0,02 °C at –5°C (±0.03 °F at 23 °F) | ±0,02 °C at –5°C (±0.03 °F at 23 °F) | ±0,02 °C at –5°C (±0.03 °F at 23 °F) | ±0,03 °C at 150 °C (±0.05 °F at 302 °F) |
Display Resolution | All temperatures 0,1 °C / 0,01 °F | All temperatures 0,1 °C/ 0,01 °F | All temperatures 0,1 °C / 0,01 °F | All temperatures 0,1 °C / 0,01 °F |
Units of measure | °C - °F - K | °C - °F - K | °C - °F - K | °C - °F - K |
Switch test | yes | yes | yes | yes |
Calibrator size | 160 x 340 x 330 mm (6,2x13,3x12,9 in.) | 160 x 340 x 330 mm (6,2x13,3x12,9 in.) | 160 x 340 x 330 mm (6,2x13,3x12,9 in.) | 160 x 340 x 330 mm (6,2x13,3x12,9 in.) |
Calibrator weight | 8 kg (17 lb) |
8 kg (17 lb) |
8 kg (17 lb) |
8 kg (17 lb) |
Power supply | 115 or 230 VAC - 50/60 Hz | 115 or 230 VAC - 50/60 Hz | 115 or 230 VAC - 50/60 Hz | 115 or 230 VAC - 50/60 Hz |
Electric power | 300 W | 300 W | 500 W | 500 W |