產品介紹∣THERMOSTATIC CALIBRATORS
Search
THERMOSTATIC CALIBRATORS

簡介

QUARTZ

  • Temperature range: -30 °C ÷ 150 °C
  • Precision: ±0,15 °C
  • Hole depth: 130 mm
  • Hole diameter: 35 mm

我要諮詢

Operating range:-30 °C / +150 °C, (-22 °F / 302 °F)

Applications:

Control and calibration of temperature sensor, in the laboratory and in the field, in conformity with ISO 9000 standards.

Control of thermostats.

Automatic computer-controlled calibrations.

Operating range --30 °C ÷ +150 °C (*) , (-22 °F / 302 °F ) (*)
Stability ±0,03 °C (at 0 °C) ± 0.05 °F (at 32 °F)
Uniformity (at 0 °C) -Radial ±0,02 °C (at 40 mm),± 0.03 °F (at 1.5 in.)
Uniformity (at 0 °C) -Axial ±0,1 °C (for 60 mm from the bottom)
± 0.1 °F (for 2.3 in. from the bottom)
Maximum heating time 20 °C/min (68 °F/min)
Maximum cooling time 22 °C/min (71.6 °F/min)
Display resolution All temperatures: 0,1 °C / 0,01 °F
Display accuracy ±0,15 °C (± 0.27 °F)
Units of measurement °C - °F - K
Serial interface RS232
Well diameter 35 mm (1.3 in.)
Insert hole depth 130 mm (5.1 in.)
Power supply 115 or 230 VAC - 50/60 Hz
Electric power 300 W
Calibrator size 160 x 340 x 330 mm, (6.2 x 13.3 x 12.9 in.)
Calibrator weight 8 kg (17 lb)
Shipping weight with carrying case 17 kg (37 lb)

(*) Room temperature 20 °C