Digital pressure gauges of the TANTUM series are particularly suitable for both stationary and mobile measurement and display of pressures. The use as a pressure reference allows easy checking, setting and calibration of other pressure measuring devices directly on site
Technical data | |||
Type | E2 | D2 | C2 |
Pressure ranges | Between -1...1000 bar | ||
Accuracy (± of full scale value) | 0.5 % | 0.1 % | 0.05 % (0.1 % in vacuum) |
Units | |||
Pressure | bar / mbar, kPa / MPa, PSI, kg/cm², mH2O, inH2O | ||
Display | |||
Type | Multifunctional LCD, 4 ½ digit, with background lighting | ||
Functions | Bar graph, pressure unit, display filter, MIN/MAX value, Overpressure indicaton 110 % of full scale pressure value (FS) |
||
Measuring system | |||
Pressure measuring cell | ≤ 10 bar piezo resistive ≥ 20 bar thin film strain gauge |
||
Temperature compensation | 0...50 °C | ||
Temperature error | 0.005 % of full scale pressure value (FS) / K | ||
Measuring rate actual value | 10 ms | ||
Measuring rate bar graph | 50 ms | ||
Process connection | |||
Thread ISO 1179-2 | G¼ with NBR gasket, AF27 | ||
Material | Stainless steel 1.4404 | ||
Medium temperature | -20...80 °C | ||
Housing connection | 270° axially rotatable | ||
Housing | |||
Degree of protection DIN EN 60529 | IP67 | ||
Dimensions | Ø 90 mm, D=45 mm H=120 mm with protection cap | ||
Materials | Cast zinc with protection cap made of TPE | ||
Ambient conditions | 0...50 °C with air humidity ≤85 % r.h. | ||
Weight | 540 g | ||
Vibration resistance IEC 60068-2-6 | 5g up to 500 Hz | ||
Shock resistance IEC 60068-2-29 | 25g for 11 ms | ||
Power supply | |||
Auto-off function | Selectable switch-off time 5 minutes | ||
Battery type | 2x 1.5 V AA | ||
Battery operation | 1500 h | ||
Data logger (optional) | |||
Max. data sets | 27000 per recording series | ||
Record series | 1...10 | ||
Storage rate | 0.1…6.4 s - automatically per recording duration | ||
Max. recording duration | 48 h | ||
Data set information | time / actual pressure / MIN/MAX value | ||
Digital data transmission | serial USB 2.0 | ||
Interface | Mini USB port | ||
PC software | Detection as USB data storage device, spreadsheet program | ||
File format | .csv | ||
Adjustment possibility (optional) | |||
Linearization | The factory linearization is changed via OFFSET and INCREASE. A special programming adapter is required. Communication between PC/laptop and digital pressure gauge is via TTL level by means of a PC terminal program |