簡介
Nuclear
這款獨特的火焰光度計是我們與德國的傑出經銷商 Wolfgang Glock 先生共同合作設計,具有 4 個用於鋰檢測的主動通道。每個通道都在檢測限範圍內獨立校準,從而提高了準確性,以滿足行業需求。
我要諮詢這款獨特的火焰光度計是我們與德國的傑出經銷商 Wolfgang Glock 先生共同合作設計,具有 4 個用於鋰檢測的主動通道。每個通道都在檢測限範圍內獨立校準,從而提高了準確性,以滿足行業需求。
每個偵檢測通道皆可同步偵測與顯示,並可透過兩線式類比輸出( 4-20m A)與SCADA™系統連接,使裝置可進行線上資料採集與監控。
透過我們的 FP-PC 應用程式輸出數據,可以透過互聯網或內部網路共享數據,以便用戶共享資訊。
客戶可以利用我們獨特的選配功能「Collection Cup」,實現線上即時的分析以及連續測量。兩線式類比輸出連接,可以將 BWB NUCLEAR FES 升級為「線上」、即時、全天運作的分析儀。藉由兩線式類比輸出( 4-20m A),BWB FP 可以與任何 SCADA™ 系統無縫整合。
產品特點:
Sample rate 2 – 3.5ml/min
Tubing materials Silicone and Tygon®
Required desk space for Instrument 50cm (H) x 45cm (W) x 45cm (D)* For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size Nuclear Flame Photometer
Shipping
Weight Nuclear Flame Photometer – 14kg (30.8lbs) Shipping – 25kg (55.1lbs) |
Optimal range Single point calibration Li 0.05 – 50ppm
Multi Point Calibration Li 0.05 – 1000ppm
Reproducibility <1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Chemical Interference Na/K/Li = <0.5% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and LOD LOQ Li – 0.02ppm Li – 0.05ppm
Time to stability Less than 15 seconds after sample is introduced into the flame |
Ions measured 4 independent Li channels, each providing up to 10 points of calibration
Interfaces USB 0-2.5 volt output (based on sample concentration linked to element of users choice) Optional 4-20mA output in place of the above Optional integrated printer .csv and .pdf generated reports and files via FP-PC software
Recommended minimum warm up 21ºC – 25 minutes
Power requirements 100V – 250V AC at 50 or 60Hz automatically selecting
Fuel requirement Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min *with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout LCD, four line, alpha numeric, back lit. |
Sample rate 2 – 3.5ml/min
Tubing materials Silicone and Tygon®
Required desk space for Instrument 50cm (H) x 45cm (W) x 45cm (D)* For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size Nuclear Flame Photometer
Shipping
Weight Nuclear Flame Photometer – 14kg (30.8lbs) Shipping – 25kg (55.1lbs) |
Optimal range Single point calibration Li 0.05 – 50ppm
Multi Point Calibration Li 0.05 – 1000ppm
Reproducibility <1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Chemical Interference Na/K/Li = <0.5% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and LOD LOQ Li – 0.02ppm Li – 0.05ppm
Time to stability Less than 15 seconds after sample is introduced into the flame |
Ions measured 4 independent Li channels, each providing up to 10 points of calibration
Interfaces USB 0-2.5 volt output (based on sample concentration linked to element of users choice) Optional 4-20mA output in place of the above Optional integrated printer .csv and .pdf generated reports and files via FP-PC software
Recommended minimum warm up 21ºC – 25 minutes
Power requirements 100V – 250V AC at 50 or 60Hz automatically selecting
Fuel requirement Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min *with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout LCD, four line, alpha numeric, back lit. |