PM1A:0~600V AC/DC,精度:±3% rdg. ±2位數;CAT. III - 600V。
PM5A2:10~1000V AC/DC,精度:±(5% rdg.+3位數);CAT. III - 600V。
PM10A:15~600V AC/DC,精度:±(5% rdg.+3位數);CAT. III - 600V。
PM15-4A:不適用。
輸入:
PM1A:內部可充電電池。
PM5A2:內部可充電電池。
PM10A:內部可充電電池。
PM15-4A:內部可充電電池或輸入電源100-240VAC,50/60Hz。
詳細規格、配件、軟體及選購,歡迎洽詢。
Phenix Technologies’ line of Insulation Analyzers perform multiple testing functions including:
Insulation Resistance
Dielectric Absorption Ratio
Polarization Index
Capacitance
Leakage Current
Guarded Current measurements
Step Voltage
Pass/Fail test
Low Resistance and Continuity measurement
AC/DC Voltmeter up to 600 V True RMS
Design Features:
Perform multiple testing functions
Simple set-up and operation
Reliable and accurate measurements in the laboratory or in the field
Automated testing with microprocessor controller
Auto-range
External noise rejection
Rechargeable battery
IEC compliant
Model PM1A Model PM5A2 and PM10A:
Insulation Resistance testing up to 200 GΩ
Pass / Fail test
Low Resistance and Continuity measurement
AC/DC Voltmeter up to 600 V True RMS
Automatic Polarization Index
Model PM5A2 and PM10A:
Insulation Resistance testing up to 5 TΩ (PM5A2) or up to 10 TΩ (PM10A)
AC/DC Voltmeter up to 1000 V (PM5A2) or 600 V (PM10A) True RMS
Pass / Fail test
Automated tests:
Absorption Index
Polarization Index
Capacitance
Leakage Current
Step Voltage
Model PM15-4A:
Insulation Resistance testing up to 15 TΩ
Automated tests:
Dielectric Absorption Ratio
Polarization Index
Capacitance
Insulation Resistance
Step Voltage
Test Voltages & Accuracy:
PM1A: 250 V - 500 V - 1000 V, DC, negative, -0 / +15% for resistances between 10 MΩ and open circuit
PM5A2: 500 V – 1 kV - 2.5 kV – 5 kV directly, one button selectable; 500 V to 5 kV in 25 V, 100 V or 500 V steps; DC, negative; ±3% of nominal test voltages on 10 GΩ
PM10A: 500 V – 1 kV - 5 kV – 10 kV directly, one button selectable; 500 V to 10 kV in 25 V, 100 V or 500 V steps; DC, negative; ±3% of nominal test voltages on 10 GΩ
PM15-4A: 0-15 kV,keypad, operator selectable; 100 V to 15 kV in 10 V steps; DC, negative; ±0.8% of Reading, ±0.2% of range, > 100 VDC
Resistance Measurement & Accuracy:
PM1A: 10 kΩ to 200 GΩ, ±5% of reading ±2 digits
PM5A2: maximum 5 TΩ@5 kV
PM10A: maximum 10 TΩ@10 kV
PM15-4A: 125 kΩ to 15 TΩ;±5% of Reading up to 2 TΩ, ±20% of Reading up to 15 TΩ
Leakage Current Measurement & Accuracy:
PM1A: maximum 1.5 mA, ±0.3 mA
PM5A2: maximum 3 mA, ±(10% of reading + 3 digits)
PM10A: maximum 1.5 mA, ±(10% of reading + 3 digits)
PM15-4A: maximum 4 mA, ±0.8% of reading, ±0.2%
Capacitance Measurement & accuracy:
PM1A: N/A
PM5A2: 50 nF up to 10 μF @ 500 V; 50 nF up to 5 μF @ 1000 V; 30 nF up to 2 μF @ 2500 V; 30 nF up to 1 μF @ 5000 V; ±10% of reading ± 3 digits
PM10A: 50 nF up to 10 μF @ 500 V; 50 nF up to 5 μF @ 1000 V; 30 nF up to 2 μF @ 2500 V; 30 nF up to 1 μF @ 5000 V; 30 nF up to 680 nF @ 10,000 V; ±10% of reading ± 3 digits
PM15-4A: 6 nF up to 100 μF @ 5000 V; ±5% of reading ± 5 nF