Stand alone unit (Standard Capacitor must be purchased separately)
Built-in RS232 interface and Windows software package included
Easily display and process precise measurement results
Precision x 10-5 measurements
Automatic balancing
Testing Applications:
The Model TG-3 is a fully automatic capacitance and dissipation factor (power factor) measuring bridge for high precision high voltage insulation diagnostics.
The TG-3 is used in applications where it is necessary to determine the tangent delta of liquid or solid dielectrics.
Maximum current through test object: 5 A (optional current extension available up to 20 A)
Maximum current through standard capacitor: 30 mA
Capacitance measuring range: 0.1 x CN ... 1,000 x CN
Dissipation factor measuring range: 1 x 10-5 ... 1
Voltage measurement: 10 V ... 1 MV
Frequency measurement: 45 Hz … 65 Hz
Uncertainty of capacitance measurement: ± 0.01%
Uncertainty of phase measurement: ± 0.02 mrad ± 1% of the displayed value
Uncertainty of voltage measurement: ± 1% of range ± 2 digits
Reference Conditions:
Minimal current through standard capacitor: 35 μA
Temperature 23°C ± 2°C
Humidity 45 ... 75%
Rated Range of Use:
According IEC 359
Temperature 5°C ... 40°C
Humidity, non-condensing 20 ... 95%
Please feel free to contact for detailed specifications, accessories, software and options.