The ProboStat™ system can be equipped with a liquid cell accessories to measure the dielectric constant of liquids at elevated temperature.

Typical use:

  • Dielectric measurements of liquids at RT - 200°C
  • Replaces a standard sample support tube assembly 
  • Use in combination with e.g. heating mantle 
  • ca. 0.1 – 0.5 ml measured volumes

我要諮詢

The current version can be used up to about 200°C, but the system can be modified to be used at even higher temperatures.

The liquid cell accessories comprise the following special parts:

  • A large-diameter metal support tube on a socket that fits on the pedestal of the ProboStat™.
  • A metal sample container cup that forms the sample compartment walls and acts as guard electrode. The bottom electrode is fastened in the cup by a PEEK insulator and Viton O-ring seals. These parts are detachable for replacement and rinsing.
  • Metal top electrode, centering plate.
  • Inner and outer electrode contacts, thermocouple.
  • Teflon sample compartment rings with a groove ring to define the sample volume. The sample volume is 16 mm in diameter and 0.50, 1.00 or 2.00 mm high. Teflon cylinder with cut-outs for spring load and electrode wires.
  • Spring load assembly consisting of insulated Kanthal wire.
  • The materials used that are in contact with the liquid are brass, PEEK, and Teflon.

Electrical specifications:

  • Geometrical factors (area/distance): 0.4,  0.2, and  0.1 m
  • Parasitic capacitances (typical): 0.3 - 0.4 pF.

The current version can be used up to about 200°C, but the system can be modified to be used at even higher temperatures.

The liquid cell accessories comprise the following special parts:

  • A large-diameter metal support tube on a socket that fits on the pedestal of the ProboStat™.
  • A metal sample container cup that forms the sample compartment walls and acts as guard electrode. The bottom electrode is fastened in the cup by a PEEK insulator and Viton O-ring seals. These parts are detachable for replacement and rinsing.
  • Metal top electrode, centering plate.
  • Inner and outer electrode contacts, thermocouple.
  • Teflon sample compartment rings with a groove ring to define the sample volume. The sample volume is 16 mm in diameter and 0.50, 1.00 or 2.00 mm high. Teflon cylinder with cut-outs for spring load and electrode wires.
  • Spring load assembly consisting of insulated Kanthal wire.
  • The materials used that are in contact with the liquid are brass, PEEK, and Teflon.

Electrical specifications:

  • Geometrical factors (area/distance): 0.4,  0.2, and  0.1 m
  • Parasitic capacitances (typical): 0.3 - 0.4 pF.